Your purchase will NOT include: - other accessories not shown or listed here.This Autoclave is from an open box that had incurred damage to the box during shipping. The sterilizer was removed from the packing and inspected, and one of the door mounts on the plastic door panel was found to be broken. The breakage was easily repairable and was repaired using epoxy cement. Then to insure the autoclaves functionality, it was tested by running sterilizing cycles of 250 F for 30 minutes and 270 F for 15 minutes. To verify that this sterilizer is achieving it's maximum required sterilization temperature during a cycle, it was tested using a Max Register Thermometer, and found to exceed 270 F for an entire cycle
Technician Testing Notes
This Autoclave is from an open box that had incurred damage to the box during shipping. The sterilizer was removed from the packing and inspected, and one of the door mounts on the plastic door panel was found to be broken. The breakage was easily repairable and was repaired using epoxy cement. Then to insure the autoclaves functionality, it was tested by running sterilizing cycles of 250 F for 30 minutes and 270 F for 15 minutes. To verify that this sterilizer is achieving it's maximum required sterilization temperature during a cycle, it was tested using a Max Register Thermometer, and found to exceed 270 F for an entire cycle
This Autoclave is from an open box that had incurred damage to the box during shipping. The sterilizer was removed from the packing and inspected, and one of the door mounts on the plastic door panel was found to be broken. The breakage was easily repairable and was repaired using epoxy cement. Then to insure the autoclaves functionality, it was tested by running sterilizing cycles of 250 F for 30 minutes and 270 F for 15 minutes. To verify that this sterilizer is achieving it's maximum required sterilization temperature during a cycle, it was tested using a Max Register Thermometer, and found to exceed 270 F for an entire cycle
Product Condition
- Manufacturer: Tuttnauer
- Model: 1730MKV ValueKlave
- Serial Number: 22080353SNM
- Power Requirements: 120V
- Year: Not available
- Sterilizer Type: steam chamber
- Timer Function: features a basic timer with one timing option
- Controls: analog control knobs
- Chamber Depth: 30 cm
- Chamber Diameter: 17 cm
- Number of Trays: 3 trays
- Cycle Count: 2
Item Includes:
1 - Tuttnauer 1730MKV ValueKlave 3 - Sterilizing Trays w/ Rack and Tray Handle 1 - Operations and Maintanence Manual Package 1 - Power Cord 1 - Original Shipping Container
1 - Tuttnauer 1730MKV ValueKlave 3 - Sterilizing Trays w/ Rack and Tray Handle 1 - Operations and Maintanence Manual Package 1 - Power Cord 1 - Original Shipping Container
Shipping calculated at checkout.
30 Day Return Guarantee
Your satisfaction is important to us: Have your equipment installed by a factory-trained technician or Atlas-approved vendor to qualify for our return policy.See our Terms and Conditions of Sale for details.